Page 126 - Intersurgical Product Catalogue Edition14
P. 126

126
Nebuliser Therapy • Nebulisers






Choosing a nebuliser


Particles above 5–10 microns impact 
Particles smaller than 0.5 microns on the upper respiratory tract
expired back to atmosphere


Nasal cavity




Pharynx 


Oral cavity
Larynx 



Trachea



Micro CirrusM HOT TopM2 CirrusM2 
alveolar deposition Tracheobronchial Tracheobronchial 

(0.5–2μm)
deposition (2–5μm) deposition (2–5μm)
high speed nebulisation 

for greater patient 
compliance








10 8 86
10 64
2
42





Tracheobronchial deposition
Alveolar deposition








10 10 88
66 44
22


Micro Cirrus nebuliser
HOT Top2 nebuliser
Cirrus2 nebuliser


The Micro Cirrus has been The HOT Top2 is designed for high The Cirrus2 has been designed to 
speciically designed to create a speed nebulisation. It signiicantly minimise the residual volume and 

small particle aerosol with a MMD reduces treatment time compared reduce drug wastage.
of 1.2 microns.
with conventional nebulisers.
At a driving gas low of 8 L/min, 74% 

At a driving gas low of 8 L/min At a driving gas low of 8 L/min, 80% of the volume output will be particles 
90% of the particles will be less of the volume output will be particles less than 5 microns in diameter with 

than 2 microns in diameter.
of less than 5 microns in diameter a MMD of 3.3 microns.
with a Mass Median Diameter 

(MMD) of 3 microns.













Intersurgical • Quality, innovation and choice

   124   125   126   127   128